Product Use

  1. Wash your hands and remove the Humidifier Bottle from packaging.
  2. Use sterile, distilled or boiled tap water. If using boiled tap water, cool water to room temperature before adding to the Humidifier Bottle.
  3. Unscrew the lid and add water to the maximum level marking located on the side of the bottle.
  4. Replace the lid. Ensure that the threads are properly aligned and that the lid is screwed on tight.
  5. Attach the Humidifier Bottle to an oxygen outlet.
  6. Test for leaks. Turn the flow to 3 LPM and block the oxygen outlet port on the  Humidifier Bottle. A sound should emit within 5 seconds. If there is no sound, check the lid and other connections for leaks. Repeat the leak test.
  7. Attach the oxygen supply tubing to the oxygen outlet port  and adjust oxygen flow to the prescribed setting.
  8. Check for bubbles in the water. If there are no bubbles check for leaks. If the device sounds, check for tubing obstruction.
  9. Apply your nasal cannula or oxygen mask.
  10. If water level drops below the minimum level marking on the Bubble Humidifier add more water.
  11. Empty and replace the water in Humidifier Bottle daily.

 

Cleaning Instructions

  1. Every 14 days wash the  Humidifier Bottle with warm soapy water. Use a non-scented, lotion free liquid dish detergent. If you use tap water in your Humidifier Bottle , clean the Humidifier Bottle every 3 days to help prevent buildup of mineral deposits. Do not use bleach or ammonia solutions to clean your device.
  2. Rinse the Humidifier Bottle thoroughly with clean water.
  3. After cleaning the device, visually inspect to ensure that it is clean. Repeat cleaning if necessary.
  4. Replace the Humidifier Bottle every 90 days (3 months), or sooner if unable to remove mineral deposits by cleaning.

Features

  1. Latex free
  2. 360 degrees molded, low resistance, high output diffuser head design
  3. Audible 6psi pop off alarm
  4. Jar permanently etched with large print minimum and maximum water levels